Olympus/Panametrics 25DLHP Thickness Gauge

Press On/Off Key The unit will turn on
Select which transducer you are going to use.
Three are supplied with the kit.
  1. M102 General Purpose
  2. M106 Thick Composites and Fibreglass
  3. M2008 Thiner Composites and Fibreglass
You may need to experiment with each probe to determine which is the most appropriate.
To configure the unit you will need two test pieces of the same material which you want to test that each have known thicknesses. One thick and one thin.
Press 2nd F  
Press CAL (Setup) A list of transducers will appear
Use the Up and Down arrow keys and select the transducer which you have decided to use
Press Enter  
Press MEAS  
  • If you are using the M2008 transducer hold the transducer in the air and press Zero briefly
  • Place couplant on the thick material.
  • Hold the transducer to the thick test piece
  • Wait for a stable reading
Press CAL(Setup) If locked appears press MEAS then 6&3 together to unlock
Press VEL CAL & VEL will appear on the display
  • Enter the known thickness of the thick test piece using the keypad
  • Place couplant on the thin material.
  • Hold the transducer to the thin test piece
  • Wait for a stable reading
Press CAL(Setup)  
Press ZERO CAL & ZERO will appear on the display
  • Enter the known thickness of the thick test piece using the keypad
Press MEAS This will save the calibration figure